96-06434-00-01 revA
Service Manual, Inogen One G3HF
- 122 -
3. In order to remove the receptacle, the corresponding column must be removed. Locate
the column latch on the underside of the device (Figure 4) and push downward on it
while simultaneously pulling the column out of the chassis. Remove the column and set
aside.
Figure 4: Location of column latch and direction to pull column
4. Remove the receptacle filter disc by blowing it out of the receptacle with compressed air.
Set aside for later reinstallation.
5. Using a T10 torx driver, unscrew the receptacle from the underside of the chassis while
pressing down on the column latch to lower the column latch teeth (Figure 5). Remove
the receptacle through the bottom of the device, slowly pulling the tubing through the
chassis. The receptacle washer should remain in place on the other side of the chassis.
Figure 5: Push down on column latch while unscrewing receptacle
Latch