![Inogen One G3HF Service Manual Download Page 112](http://html1.mh-extra.com/html/inogen/one-g3hf/one-g3hf_service-manual_2064301112.webp)
96-06434-00-01 revA
Service Manual, Inogen One G3HF
-112-
7.
Slide the filter body, filter, compressor, and dual feed tube out of the device as a single
unit (Figure 7).
Figure 7: Removal of compressor assembly
8.
Remove one end of the dual feed tube off of the compressor at a time (Figure 8) by
pushing the tube down between the barbs.
Be careful not to damage the sealing area of
the barb during removal.
Figure 8: Remove dual feed tube from compressor
Instructions for Replacing the Dual Feed Tube:
1. Use the same technique to install the new dual feed tube to the compressor. Use pliers to
push the tube onto the connectors (Figure 9).