96-06434-00-01 revA
Service Manual, Inogen One G3HF
-113-
Figure 9: Properly installed dual feed tube
1.
Slide the compressor, filter body, filter, and new dual feed tube back into the device as a
single unit. The filter body slides into specific grooves in the chassis (Figure 10). Use these
grooves to guide the whole compressor assembly into the device.
Figure 10: Filter body grooves
3. Reattach the two screws on the front of the device and the bottom center screw on the
back of the device that connect the filter body to the chassis (Figures 5 and 6). Tighten
screws to 6 in-lbs and 4.5 in-lbs respectively.
4. Reinstall the screw and washer that connect the dual end of the feed tube to the chassis,
tightening to 3 in-lbs. (Figure 3).