![Inogen One G3HF Service Manual Download Page 47](http://html1.mh-extra.com/html/inogen/one-g3hf/one-g3hf_service-manual_2064301047.webp)
96-06434-00-01 revA
Service Manual, Inogen One G3HF
-47-
2. Place the device on its back side and remove the two screws that secure the front
housing to the bottom of the chassis (Figure 2).
Figure 2: Location of screws
3. With the concentrator still on its back side, slide the front housing down and away
from the device then lift up to remove.
Instructions for Removing the Compressor:
1. Locate and remove the two screws that connect the tube bracket to the chassis (Figure 3).
Remove the tube bracket and set aside.
Figure 3: Tube bracket screws
2. Remove the four screws attaching the filter box lid to the filter body (Figure 4) and
remove the filter box lid using a flathead screwdriver to separate it from the filter body,
if necessary.