![Inogen One G3HF Service Manual Download Page 104](http://html1.mh-extra.com/html/inogen/one-g3hf/one-g3hf_service-manual_2064301104.webp)
96-06434-00-01 revA
Service Manual, Inogen One G3HF
-104-
Instructions for Replacing the Tube Bracket:
1.
Install the new tube bracket, making sure that the pegs and intake tube are in their
respective slots and grooves on the bracket (as seen in Figure 3) before replacing the two
screws. Tighten screws to 4 in- lbs.
Instructions for Replacing the Housing:
1. Place the device on its back side.
2. Seat the feet of the front housing into the corresponding grooves on the chassis (Figure 4)
then push down and toward the device to secure the housing in the chassis.
Figure 4: Housing feet and corresponding chassis grooves
3. Replace the two screws that connect the front housing to the chassis (Figure 2). Tighten
screws to 3 in-lbs.
4. Return the concentrator to the upright position
5. Slide the battery back into the chassis. Ensure the battery clip is fully seated.