Ecotron - Incubation Shaker
Installation
03 May 2018
Page
41
of
90
6 Installation
6.1 Requirements for the Installation Site
6.1.1 Operating Conditions at the Installation Location
To achieve optimal and reproducible results, the equipment should
be set up in stable ambient conditions without strong temperature
or humidity fluctuations.
All climatic changes in the environment can transfer to the climate
in the interior, after which the control system will require some time
to reach the setpoints again.
Operating conditions
Temperature range
10 – 30°C
Humidity
10 – 85%
Restrictions
Do not expose to direct sun-
light
Do not expose to dust
Do not expose to vibrations
The equipment may only be set up indoors.
To achieve a stable climate in the incubation chamber, the follow-
ing requirements must be observed:
Do NOT set up directly underneath, next to or in the air stream
of the air conditioners or other devices that generate strong air
flows (e.g. circulation fans in clean rooms).
Do NOT set up next to equipment with strong heat radiation,
such as heaters or autoclaves, and do not place in their ex-
haust heat stream.
Do NOT set up next to equipment with strong heat radiation,
such as heaters or autoclaves, and do not place in their ex-
haust heat air stream.
INFORMATION
In order to ensure that the climate in the incubation chamber
can be controlled as precisely as possible, the equipment must
not be set up in poorly ventilated recesses or in the exhaust
heat air stream of air conditioners or other sources of heat.