Ecotron - Incubation Shaker
Spare Parts
03 May 2018
Page
73
of
90
10.2 Replacing Control Rack
If the electronics fail (control of the shaker drive and tempering) or
if there is a defect in the operating panel, power supply unit or fan,
only one component has to be replaced, the control rack. To do so,
proceed as follows:
CAUTION
The round heating element as well as the lighting element in the
control rack can heat up significantly during operation. Hence,
you could get burnt if you touch the round heating element or
lighting elements.
– Before removing the control rack, let the round heating ele-
ment and the lighting elements cool down to ambient tem-
perature.
Procedure
1.
Turn off equipment and pull out the plug.
2.
Open the door to the incubation chamber.
3.
Remove the deflector plate on the ceiling of the incubation
chamber by loosening the two Phillips-head screws (M4x10).
4.
Detach the cable from the shaker to the control rack and pull it
out of the cable duct. To detach the plug, you have to push
the release lever at the bottom of the plug.