Ecotron - Incubation Shaker
Operation
Page
56
of
90
03 May 2018
7.4 Adjusting, Activating and Deactivating Parameters
coi
7.4.1 Overview of Parameters
Rotation speed (
RPM
)
During set-up, the rotation speed entered as a setpoint in min
-1
is
displayed as a whole number. The settings that can be made range
from 20 to 550 min
-1
.
The maximum permissible rotation speed depends on the throw,
the position of the equipment in a stack and the load on the tray
(mass) (see chapter 12.4.1 "Maximum Permissible Setpoints for
the Rotation Speed", page 85).
Temperature (
°C
)
During set-up, the set temperature setpoint in degrees Celsius is
displayed to one decimal point. The setting range is between 0 and
65°C.
The temperatures that can actually be achieved depend on a vari-
ety of factors, such as the ambient temperature around the equip-
ment, the ventilation and the temperature of the other equipment in
the stack (see chapter 12.4.2 "Information on Temperatures that
can Actually be Reached", page 86).
INFORMATION
To protect the compressor from overloading, cooling switches
on with a time delay of up to three minutes in certain situations.