Ecotron - Incubation Shaker
Spare Parts
03 May 2018
Page
75
of
90
10.4 Replacing the Drip Tray
If the drip tray gets very dirty or damaged, it can be removed and
replaced. To do so, proceed as follows:
INFORMATION
The drip tray is not suitable for steam sterilisation in the auto-
clave. If it gets severely contaminated it can be disposed of and
replaced at low cost.
Procedure
1.
Turn off equipment and pull out the plug.
2.
Open the door to the incubation chamber.
3.
If a tray is installed, remove it by loosening the four Phillips-
head screws (M4x6).
4.
Remove the table by loosening the four hexagonal socket
screws (M6x12).
5.
Drain any possible liquid from the drip tray.
6.
First, lift the drip tray diagonally up to 45°, then lift the front un-
til the tray can be moved back via the guide bearings.
7.
Insert the new drip tray and install the table in reverse order.