15 Troubleshooting
15.1 Faulty Symptoms
Fault Symptom
Fault
Measures
No sound in Detection
Mode and Analysis Mode.
Sound set low.
Click the speaker icon to
increase the volume
Broken speaker or speaker
cable loose.
Send in for repair.
No picture on the display,
no sound.
Fuse broken.
Check fuse. Replace if
broken.
Main Cable broken.
Replace the Main Cable.
Discharged battery
(Portable Model).
Charge the battery.
Main PCB broken.
Send in for repair.
No picture on display
Display broken or lamp
cable loose.
Send in for repair.
White picture on display.
Main PCB broken or
display cable loose.
Send in for repair.
No working display touch
function
Display broken or touch
cable loose.
Send in for repair.
Loss of spots or black lines
on screen.
Display broken.
Send in for repair.
Bad sound out of the
speaker.
Broken speaker.
Send in for repair.
Short operation time on
Portable Model.
Old or broken battery.
Send in for repair.
The time and dates resets
at restart.
The clock battery is empty. Send in for repair.
Settings lost on restart
Main PCB Memories
broken (Flash Memories).
Send in for repair.
Low or no gas sensitivity.
Old or broken gas sensor.
Probe broken.
Replace sensor.
No prob lamp light
Broken lamp.
Broken probe PCB.
Send in for repair.
Send in for repair.
No display light or partly
picture
No display light or partly
picture.
Send in for repair.