Parameter
Read
Write
Sequence
of Data
Answer
Sequence
of Data
value/text
MEAS_GAS
rP21\n
Displayed Gas:
[text]
wP21 [text]\n
If write 95%N
2
- 5%H
2
,
Air, He, N
2
, H
2
, R123a,
R22, R290, R404a,
R407c, R410, R600a,
R1234yf automatic
converting is enabled.
Max 14 characters.
MEAS_GAS_VISCOSITY
rP22\n
Displayed Gas
Viscosity (µPas):
[value]
wP22 [value]\n
Enable when Displayed
Gas not set to: 95%N
2
-
5%H
2
, Air, He, N
2
, H
2
,
R123a, R22, R290,
R404a, R407c, R410,
R600a, R1234yf
MEAS_GAS_DENSITY
rP23\n
Displayed Gas
Density (g/l):
[value]
wP23 [value]\n
Enable when Displayed
Gas not set to: 95%N
2
-
5%H
2
, Air, He, N
2
, H
2
,
R123a, R22, R290,
R404a, R407c, R410,
R600a, R1234yf
MEAS_GAS_SHOW
rP24\n
Displayed Gas
Name: [text]
wP24 [text]\n
0 = On
1 = Off
REJECT_LEVEL
rP25\n
Reject Level:
[value]
wP25 [value]\n
REJ_CHOPPED_AUDIO
rP26\n
Chopped Audio
Signal: [text]
wP26 [text]\n
0 = On
1 = Off
REJ_PROBE_FLASH
rP27\n
Probe Lamp Flash:
[text]
wP27 [text]\n
0 = On
1 = Off
SHOW_REJECT_LEVEL
rP28\n
Show Reject
Level: [text]
wP28 [text]\n
0 = On
1 = Off
MIN_PRESENT_TIME
rP29\n
Min Presentation
Time (s): [value]
wP29 [value]\n
0.1 – 100
(1=0.1 s)
DISPLAY_THRESHOLD
rP30\n
Display Threshold
(%): [value]
wP30 [value]\n
0 – 99
MEAS_AUDIO_THRESHOL
D
rP31\n
Measuring Audio
Threshold (%):
[value]
wP31 [value]\n
0 – 99