14 | Serial Communication
INFICON
46 / 80
Sensistor_Sentrac_Strix_Operating_Manual_ninb65en1-01-(2010)
Parameter
Read
Write
Sequence
of Data
Answer
Sequence
of Data
value/text
LEAK_GAS_VISCOSITY
rP07\n
Leak Gas Viscosity
(µPas): [value]
wP07 [value]\n
Enable when Leak gas
not set to:
95%N
2
/5%H
2
, Air, He,
N
2
, H
2
, R123a, R22,
R290, R404a, R407c,
R410, R600a, R1234yf
LEAK_GAS_DENSITY
rP08\n
Leak Gas Density
(g/l): [value]
wP08 [value]\n
Enable when Leak gas
not set to: 95%N
2
- /5%H
2
, Air, He, N
2
, H
2
,
R123a, R22, R290,
R404a, R407c, R410,
R600a, R1234yf
CAL_INTERVAL_ACTIVE
rP09\n
Interval Reminder
active: [text]
wP09 [text]\n
0 = On
1 = Off
CAL_INTERVAL_DAYS
rP10\n
Days: [value]
wP10 [value]\n
0 - 255
CAL_INTERVAL_HOURS
rP11\n
Hours: [value]
wP11 [value]\n
0 - 23
CAL_INTERVAL_MINUTES
rP12\n
Minutes: [value]
wP12 [value]\n
0 - 59
LOC_SENSITIVITY
rP13\n
Sensitivity: [value]
wP13 [value]\n
1 - 15
LOC_RANGE_CHOICE
rP14\n
Locating Range:
[text]
wP14 [text]\n
0 = Manual
1 = Auto
LOC_REJECT_INDICATION
rP15\n
Reject Indication:
[text]
wP15 [text]\n
0 = On
1 = Off
LOC_READY_PULSE
rP16\n
Locating Audio
Ready Pulse: [text]
wP16 [text]\n
0 = On
1 = Off
LOC_DIRECT_ADJUST
rP17\n
Direct Sensitivity
Adjustment: [text]
wP17 [text]\n
0 = On
1 = Off
LOC_AUDIO_THRESHOLD
rP18\n
Locating Audio
Threshold (%):
[value]
wP18 [value]\n
0 - 100
MEASURE_UNIT
rP19\n
Measuring Unit:
[text]
wP19 [text]\n
If write cc/s, Pa m
3
/s,
cc/in, SCCM, g/yr, oz/
yr, mbarl/s, mm
3
/s,
mm
3
/min automatic
converting is enabled.
Max 14 characters.
CORR_VALUE
rP20\n
Correlation Value:
[value]
wP20 [value]\n