15 | Troubleshooting
INFICON
54 / 80
Sensistor_Sentrac_Strix_Operating_Manual_ninb65en1-01-(2010)
15.2 Warning Messages
Message
Indicators
Message in the
menu warning list
Code
Probe not
connected.
Blue LED flashing.
(None)
(None)
SD-card not
inserted.
Orange measure
window at screen.
(None)
(None)
Battery discharged.
Recharge battery.
Orange measure
window at screen,
fast Blue LED
flashing.
(None)
(None)
Sensor broken or
not connected.
Change sensor or
connect sensor.
(Voltage to high,
Error E1)
Red measure
window at screen,
Blue LED flashing.
Sensor broken or
not connected.
(Voltage to high)
01
Sensor broken.
Change sensor.
(Voltage to Low,
Error E2)
Red measure
window at screen,
Blue LED flashing.
Sensor broken.
(Voltage to Low)
02
Sensor broken.
Change sensor.
(Temperature to
Low, Error E3)
Red measure
window at screen,
Blue LED flashing.
Sensor broken.
(Temperature error)
03
Probe
disconnected.
Reconnect the
probe. (Error 04)
Red measure
window at screen,
Blue LED flashing.
Probe
disconnected.
04
Low sensor signal.
Reject level to low
for sensor
sensitivity.
09