16 | Maintenance Instructions
INFICON
60 / 80
Sensistor_Sentrac_Strix_Operating_Manual_ninb65en1-01-(2010)
Assembling the Cover
1. Reinstall the cover (C). Make sure that the lock washer on the Probe Connection
Port (LEMO connector on main PCB) is still present. Important for proper
grounding. Make sure that no cables are pinched between the cover and the
chassis.
2. Reinstall and tighten the two screws (B) holding the cover (C). Tighten to 4 Nm.
3. Reinstall the Probe Connection Port nut (A).
4. Use a Spanner for notched nut and tighten the Probe Connection Port nut.
5. Check the resistance between the ground pin in the mains socket and the outside
of the probe cable connector, should be 9 Ω or lower. If not, check the lock washer
on the Probe Connection Port.