9 | Operating the Instrument
INFICON
34 / 80
Sensistor_Sentrac_Strix_Operating_Manual_ninb65en1-01-(2010)
To use I•Guide
If a fixed number of measurement points is used, then measure according to the
following steps:
1. Press the probe button, or start button on the screen, to begin the first
measurement.
2. Place the probe on the test object while the time bar moves. The instrument
registers the results.
3. You might need to wait until the next measurement. The instrument signals Wait.
4. Repeat the procedure for the next measuring point.
When all measurements are made, the sum of all leaks is shown. If the sum of all
leaks is greater or the same as the Reject Level, REJECT is shown. If the sum of all
leaks is under the Reject Level, ACCEPT is shown. And if the sum of all leaks is
greater than Reject Level before all measurements are made then REJECT is shown.
To begin a new measurement action or to stop a current measurement, press the
probe button and hold the button for a moment.
It is possible to measure or search in Combined Mode for a leak without registering
the measurement. A measurement value is only registered when the time bar is
moving.
Measuring with Dynamic Measurement Points
If a dynamic number of measurement points is selected, then do the measurement in
the following steps:
1. Press the probe button, or start button on the screen, to begin the first
measurement.
2. Place the probe near the measurement point while the time bar is moving.
3. You might need to wait until the next measurement. The instrument signals Wait.
4. Repeat the procedure for the next measuring point.
5. When you sum up all the measurements press and hold on the probe button a
short time.
To view specific measurements, press the backward and forward buttons
simultaneously.
If a measurement is to be aborted and restarted, press and hold the probe button a
few seconds. You can also click the Abort button on the screen.