INFICON
Maintenance Instructions | 16
Sensistor_Sentrac_Strix_Operating_Manual_ninb65en1-01-(2010)
57 / 80
Disassemble
1. Attach the supplied Sensor Key at the bottom of the Probe Tip Protection Cap (A)
and pull with one finger on each side of the probe tip.
2. Remove the Protection Cap (A).
Assemble
1. Install the new Probe Tip Protection Cap by pressing it over the tip.
If the probe tip filter is removed, an air gap is formed between the inner metal filter and
filter in the protection cap.
The filter in the protection cap prevents water to get into the probe if the probe comes
in contact with water.
It is recommended to always replace the protection cap with a new if exposed to
water.
16.4 Changing Probe Sensor
Part
Part no.
Qty. Consumables
Strix Sensor
590-290
1
C
A
B
A
B
C
Disassemble
1. Loosen the safety nut (A) using the Sensor Key.
2. Remove the Sensor (B) by pulling it straight out.
Assemble
1. Carefully install the new Sensor. Make sure that the Probe Pipe (C) and Sensor
connects correctly.