INFICON
Serial Communication | 14
Sensistor_Sentrac_Strix_Operating_Manual_ninb65en1-01-(2010)
51 / 80
Execute Commands
Execute command Sequence of
Data
Value/text
Explanation
Bitmap Dump
cEBD\n
-
Save a bitmap of Sentrac screen on SD-card inserted into
Sensistor Sentrac. File name: [time].bmp
Operation Mode
change
cEom
[value]\n
1 … 3
1 = Localization
2 = Measuring
3 = Combined
Set analog out
cEao
[value]\n
0 … 255
0.15 - 10.0 V, with a resolution of 8 bits between set values
of Measuring Mode
Load recipe
cElr
[value]\n
1...8
Factory reset
cEfd\n
-
Settings reset
cEsr\n
-
Set Last service
-
Date set by Sensistor Sentrac's real time clock
Wake up screen
cEwu\n
-
Read File (Data)
File data
Read
Sequence of Data
Answer
Explanation
Read Data
rFsd SENTRAC.LOG\n
SENTRAC.LOG+[File
content]
(“SENTRAC:LOG No
data saved/File I/O err” if
no saved data)
(“SD-card not inserted” if
no SD-card)
Read the content of saved measuring
data on SD-card and send it as a text
file.