The display limits for the different units of measurement are summarized in the table
below.
The lower display limit may be increased by the user. Please refer to Section
for how to edit the lower display limit.
As soon as an acoustic alarm sounds, the tip should be removed from the spot being
tested. After displaying a constant leak rate the ZERO button should be pressed
again to repeat the test. Thus, a measurement error can be prevented and the leak
can be located.
Fig. 3-6 Sniffer display when detecting a leak
Unit
Lower display limit
Upper display limit
g/a
0.05
1,000
oz/yr
0.002
1,000
ppm
0.5
100,000
mbar l/s
1x10
-7
9.9x10
-2
Pa m³/s
1x10
-8
9.9x10
-3
atm cc/s
1x10
-7
9.9x10
-2
Torr l/s
1x10
-7
9.9x10
-2
Summary of Contents for Ecotec E3000
Page 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Page 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Page 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Page 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Page 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Page 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Page 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Page 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Page 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...