Pin assignment (default)
All PLC outputs can be defined by the user from a list of commands. The commands
can be selected in the SETTINGS / INTERFACES / SELECT PLC OUTPUTS
submenu.
For details on this submenu please refer to Section
Ready
Signal is LOW as long as the Ecotec E3000 is ready for measurements.
Leak
Signal is LOW if the preset trigger level is exceeded
Error
Signal is LOW if a warning or error message is active.
7.1.4.1
Relay outputs
Technical data
Open collector output
Active = low
Pin
Default Command
4
Ready
5
not used
16
Leak
17
Error
Caution
Maximum load rating is 60 V DC / 25 V AC and 1 A per relay (resistive load).
Relay
Active = Normally open contact (NO) closed
Summary of Contents for Ecotec E3000
Page 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Page 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Page 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Page 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Page 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Page 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Page 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Page 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Page 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...