0-10
Important Safety Precautions
ki
n
a
22
eI
V
Z
.fm
Tech
ni
ca
l Ha
nd
bo
ok
(1
20
6)
Caution
The air filter should be checked for contamination at least every 6 months and
should be definitely exchanged after 2 years.
Caution
The lubricant reservoir may contain toxic substances from the pumped media.
Please dispose of lubricant reservoir as required by local regulations. A Safety Data
sheet for the lubricant is available on request.
Caution
Caution: Abrupt movements.
Abrupt movements can damage the running turbo pump.
Avoid abrupt movement and vibration of the instrument (e.g. running over cables,
door sills) during operation and up to 4 minutes after switching off since the turbo
pump can be damaged.