Maintenance
8-4
ki
n
a
22
e c
h
ap
te
r 8
.fm
te
chn
ica
l ha
nd
boo
k
(1
20
6)
4
Depending on the contamination the air filter should be cleaned or replaced as
necessary.
5
Slide in new (cleaned) air filter until fully in contact with dead stop.
6
Close air filter fastener.
7
From the main menu go to H
ISTORY
&
MAINTENANCE
/ C
ONFIRM
MAINTENANCE
and
confirm with the appropriate button on the right side of the display that the air filter
has been replaced / cleaned. The actual date and time as well as the actual number
of operating hours will be saved and a reminder to perform this maintenance task
again will be issued after another 10,000 operating hours.
Notice:
The Ecotec E3000 must be set to
ADVANCED
mode for confirming
maintenance tasks.
Fig. 8-2 Dismounting of the air filter
Fig. 8-3 Inserting the air filter
Summary of Contents for Ecotec E3000
Page 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Page 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Page 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Page 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Page 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Page 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Page 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Page 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Page 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...