Maintenance
8-16
ki
n
a
22
e c
h
ap
te
r 8
.fm
te
chn
ica
l ha
nd
boo
k
(1
20
6)
Notice:
Screw in the new reservoir with hex nut key only!
4
Re-install the ECO-Check in the Ecotec E3000.
Notice:
The ECO-Check does not fit in the Ecotec E3000 completely. There is a
little gap between the front panel of the Ecotec E3000 and the ECO-Check.
On the certificate which is delivered with the replacement gas reservoir you will find
a new serial number and a 12-digit-code containing the new leak rate and other leak
specific information. In the software menu please go to
HISTORY
&
MAINTENANCE
/
REPLACE
ECO
-
CHECK
. In the opening sub-menu phase enter the new serial number
in the first line and the 12-digit-code in the second line and press okay.
Fig. 8-18 Installing the gas reservoir
Fig. 8-19 Gap between ECO-Check and front panel
Summary of Contents for Ecotec E3000
Page 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Page 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Page 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Page 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Page 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Page 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Page 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Page 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Page 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...