6-2
Ecotec E3000 Messages
ki
n
a
22
e c
h
ap
te
r 6
.fm
te
chn
ica
l ha
nd
boo
k
(1
20
6)
Error no.
Plain text message
Possible reason
Recommendation
E1
24V of the MC50 too low
Fuse F1 on the motherboard has blown
Replace fuse F1 on motherboard*
E2
24V of sensor heating too low
Fuse F2 on the motherboard has blown
Replace fuse F2 on motherboard*
E3
24V II too low
Fuse F3 on the motherboard blown
Replace fuse F3 on motherboard*
W4
24V of the OPTION socket too low
Fuse F4 on the motherboard has blown
Replace fuse F4 on motherboard*
W5
5V of the sniffer is too low
Fuse F5 on the motherboard has blown
Replace fuse F5 on motherboard*
W6
5V of the internal Leak too low
ECO-Check electronics defective
Replace ECO-Check reference leak
or call nearest INFICON service
representative!
E7
-15V of the MC50 is too low
Motherboard defective
Call nearest INFICON service
representative!
E8
15V of the MC50 is too low
Motherboard defective
Call nearest INFICON service
representative!
W12
TMP frequency not reached during
start-up!
Frequency converter faulty
Turbo molecular pump faulty
If W12 is issued in a high humidity
environment or after extended
shutdown times, please shut off the
Ecotec E3000 and turn it back on.If
W12 occurs again, please let the
Ecotec E3000
sit in idle (with the
pumps running) for about 30 min.
W14
Filter service interval expired!
Time since last filter change longer than
set interval
Please replace sniffer tip filter! (see
Section
)
W16
TMP service interval expired!
More than three years since last TMP
maintenance
Please replace lubricant reservoir!
(see Section
)
W17
Fore pump service interval expired!
> 10,000 operating hours since last fore
pump service
Replace membrane of diaphragm
pump!*
W18
Air filter service interval expired!
> 10,000 operating hours since last main
air filter service
Clean or replace main air filter!
E20
Temperature at electronic unit is too
high (>60°)
Ambient temperature too high
Cool down environment, place
Ecotec E3000 in cooler area.
Ventilation failure
Check if fans on both side of main
unit are running (check for air flow
through inlets on both side of main
unit housing)
Air filter dirty
Clean or replace main air filter
E22
TMP frequency too low!
Frequency converter faulty
Tubro melecular pump faulty
Call nearest INFICON service
representative!
E23
TMP frequency too high!
Frequency converter faulty
Turbo molecular pump faulty
Call nearest INFICON service
representative!
W24
24V for the ext. control unit too low
For Ecotec E3000RC only:
Fuse on the RC-driver board blown
Replace fuse on RC-driver board!*
Summary of Contents for Ecotec E3000
Page 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Page 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Page 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Page 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Page 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Page 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Page 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Page 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Page 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...