Pin assignment (default)
All PLC outputs can be defined by the user from a list of commands. The commands
can be selected in the SETTINGS / INTERFACES / SELECT PLC OUTPUTS
submenu. For details on this submenu please refer to Section
.
Leak
If the preset trigger level is exceeded, relay contact NO will close and the relay
contact NC will open respectively.
Ready
As long as the Ecotec E3000 is ready for measurements, the relay contact NO is
closed and the relay contact NC is open respectively..
7.1.4.2
Recorder Outputs
Technical data
Pin assignment
For details please refer to section
Pin
Type of contact
Relay Output (default)
10
Currentless closed (NC)
Leak
22
Common
23
Currentless open(NO)
24
Currentless closed (NC)
Ready
11
Common
12
Currentless open (NO)
Analogue output
0 … 10 V
max. 1mA
Pin
Command
1
Leak rate, in linear or logarithmic scale
14
Selected gas number
Summary of Contents for Ecotec E3000
Page 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Page 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Page 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Page 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Page 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Page 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Page 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Page 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Page 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...