Select PLC Outputs
In the
SELECT PLC OUTPUTs
submenu the user can select which pin on the I/O port
(suitable for PLC output) represents which command. There are four PLC outputs
and two relays outputs available. The default setting is as follows:
PLC outputs
Relays outputs
To change these settings select the appropriate pin with the UP and DOWN arrows
on the left side of the display and afterwards select the desired command from the
list of commands with the UP and DOWN buttons on the right side of the display.
Press OK to save your settings. A screen with all selected settings will be displayed
for your reference. Confirm with OK again.
Pin
Command
Pin
Command
4
Ready
16
Leak
5
not used
17
Error
Pin
Type of contact
Relay Output (default)
10
Currentless closed (NC)
Leak
22
Common
23
Currentless opend (NO)
24
Currentless closed (NC)
Ready
11
Common
12
Currentless opend (NO)
Fig. 4-22 Defining PLC outputs
Summary of Contents for Ecotec E3000
Page 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Page 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Page 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Page 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Page 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Page 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Page 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Page 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Page 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...