How the Instrument Works
2-7
ki
n
a
22
e c
h
ap
te
r 2
.fm
te
chn
ica
l ha
nd
boo
k
(1
20
6)
In cases where the main unit is placed in a difficult or inconvenient to access area
the built-in ECO-Check reference leak can be removed from the main unit for easier
access and connected to the main unit via the Sub-D connector with a commonly
available extension cord. The ECO-Check reference leak can then be placed in an
area where it is convenient for the operator to be reached.
Notice:
The ECO-Check reference leak is not part of the Ecotec E3000 shipment
and needs to be ordered as a separate part no. (see section
).
Notice:
If you have not purchased the ECO-Check reference leak, warning 71 (“No
communication with test leak”) will be issued on first start-up. Please go to
SETTINGS
/
INTERFACES
/ ECO-C
HECK
and set the ECO-Check to “
DISABLED
”
to prevent future warnings (see section
)
The Ecotec E3000 must be set to
USER
MODE
„
ADVANCED
“ for access to this
submenu (See section 4.5.3). Go to
SETTINGS
/
DISPLAY
/
USER
MODE
.
Fig. 2-8 Connections for remote use
Pos. Description
Pos. Description
1
Connector at built-in test leak
2
Connector at
Ecotec E3000
housing
Summary of Contents for Ecotec E3000
Page 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Page 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Page 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Page 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Page 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Page 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Page 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Page 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Page 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...