4-38
Equipment Settings
ki
n
a
22
e c
h
ap
te
r 4
.fm
te
chn
ica
l ha
nd
boo
k
(1
20
6)
Replace ECO-Check
(in
ADVANCED
user mode only)
When replacing the gas reservoir of the ECO-Check reference leak, the new serial
number of the replacement gas reservoir as well as a code containing the calibration
data of the new reservoir needs to be entered in this submenu.
For details on how to replace the gas reservoir of the ECO-Check reference leak
please refer to Section
.
Notice:
This menu item will only be available as long as the ECO-Check is enabled
in the
SETTINGS
/
INTERFACE
/
ECO
-
CHECK
submenu.
Sniffer tip filter
(in
STANDARD
user mode only)
The maintenance interval for the sniffer tip filter depends on the operating
environment and may be set to values between 10 and 999 hours.
If set to 0 the reminder message for the sniffer tip filter will be disabled.
Default value: 100 hours
For details on how to perform the maintenance tasks see Chapter
, Maintenance.
Select Filament
In this submenu you can manually select which Transpector
TM
filament to use.
Default: A
Fig. 4-30 Setting the maintenance interval for the sniffer tip filter and confirming its
replacement
Summary of Contents for Ecotec E3000
Page 2: ...0 2 kina22eIVZ fm Technical Handbook 1206 ...
Page 13: ...General Information 1 3 kina22e chapter 1 fm technical handbook 1206 ...
Page 30: ...1 20 General Information kina22e chapter 1 fm technical handbook 1206 ...
Page 38: ...2 8 How the Instrument Works kina22e chapter 2 fm technical handbook 1206 ...
Page 58: ...3 20 Operation of the Ecotec E3000 kina22e chapter 3 fm technical handbook 1206 ...
Page 98: ...4 40 Equipment Settings kina22e chapter 4 fm technical handbook 1206 ...
Page 106: ...6 8 Ecotec E3000 Messages kina22e chapter 6 fm technical handbook 1206 ...
Page 133: ...Maintenance 8 18 kina22e chapter 8 fm technical handbook 1206 ...
Page 143: ...Annex 9 10 kina22e chapter 9 fm technical handbook 1206 9 2 Declaration of Conformity ...