Magnet Mount Kit installation
Do not use
with a unit that will draw power from a DC fixed power source.
Insert magnet post into a threaded screw hole on the back of the
case; turn clockwise to tighten.
Repeat for second magnet installation.
The two magnets may be installed in the following
configurations: top, bottom, left side, right side, and diagonal.
Two magnet mounts will support the weight of the RGX; up to
four magnets may be installed.
Intrinsic Safety (IS) Cable Adapter installation
IS Cable Adapter
connector
RGX IS port
Wide tab
Wide notch
Turn the IS power port cap
counterclockwise
(approximately 45°) to
expose the port for use.
Align the wide tab of the metal adapter connector to the wide notch on the IS port; push on to
the port to connect.
Turn the swivel connector clockwise (approximately 45°) until it clicks closed or a full-stop
connection impact is felt.
—
—
Connect the other end of the adapter in accordance with the
Product Manual
instruction from the compatible Industrial
Scientific power-supply accessory that is to be used.
Figure 5.2 Mounting kit and IS cable adapter service tasks