Key Features  

Location  

The RGX Gateway is suitable for indoor or outdoor use in locations that meet the product's 

certified uses 

and 

specifications.

 

Indoor operation is suitable only when a unit's GPS location is 

not 

needed. 

Communication and data security 

Instrument

RGX Gateway  

LENS™ Wireless (Linked Equipment Network for Safety) is a long-range, power-efficient wireless mesh 
network from Industrial Scientific. LENS functionality enables data sharing among wirelessly connected 

equipment items—gas-detection instruments and RGX Gateway units—that are set to operate within a 
specified LENS group (e.g., Group A). The following also apply.   

x

 

Ten named LENS groups are available.  

x

 

Each group can host up to 25 equipment items.  

x

 

More than one RGX can be included in a LENS group. 

Instrument data exchanged through LENS are automatically encrypted with the Industrial Scientific key. 
Encryption can also be set to use a customer-supplied key. All equipment items in the group must be set to 
the same key. 

RGX Gateway

iNet  

The RGX can send data it receives from gas-detection instruments to iNet through these transmission 

options: wireless local area network (Wi-Fi), cellular, and Ethernet.   
The customer sets each transmission option to on or off, allowing the RGX to maximize or to limit the 
number of channels through which it can exchange data with iNet. When more than one option is on, the 
customer prioritizes the order in which the unit will use them. For example, if Wi-Fi and cellular are on, the 

unit can be set to first use Wi-Fi and if that is not available, use cellular.   
The RGX–iNet data exchanges are automatically encrypted using the Industrial Scientific encryption key.  
Optionally, a customer-supplied key may be used.   

Mobility and mounting 

The versatile RGX is designed for transportability and for permanent installation. It can be left unmounted 
for more mobile applications or mounted for permanent operation.   
The RGX features a durable, factory-installed handle, which allows for a unit’s ready relocation from site to 
site when mobility is a factor. It can be transported in its optional case, which provides added protection for 
the unit, visibility to the unit's operational status indicator, and access to the unit's handle.  
Based on the application, the RGX can be surface mounted or mounted to a customer-supplied strut-type 

rail installation using the optional Wall Mount Kit. The RGX can also be secured to a metal surface using 
the optional Magnet Mount Kit; however, 

do not 

use magnet mounts with a unit that will draw power from a 

fixed DC power source. 
 

Document Number: 0000037533
DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS(S):N/A

Modified: 2018-05-25 13:18:14 EDT

State:In Work

INDUSTRIAL SCIENTIFIC CORPORATION

Title: PRODUCT MANUAL RGX GATEWAY-EN-1
Eng:

Drawing No: 17158071-1

Drawn By: Gaygan, Nancy

Created: 2018-05-16 10:24:51 EDT

Rev: 1

Printed By: Kuzmich, Bob

Int: 1.5

Date:May 29, 2018

Summary of Contents for RGX

Page 1: ...cument Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 2: ...vision 1 www indsci com rgxgateway Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 3: ...Overview 10 Chapter 2 13 Deployment Planning 13 Introduction 13 Communications 13 RGX Gateway instrument communication LENS Wireless 13 RGX Gateway iNet communication 13 Settings and Connections 14 Sample Deployment Plans 15 Chapter 3 21 Setup 21 Introduction 21 Unpack 21 Activation 22 Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 1...

Page 4: ...n Precheck 43 Chapter 5 45 Service and Warranty 45 Service 45 Supplies 45 Instruction 46 Warranty 52 Limitation of Liability 52 Appendix A 53 Supplemental information about external power connections 53 Contact Information 56 Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title P...

Page 5: ...onfiguration applications 25 Table 3 3 Supplies needed for configuration 25 Table 3 4 RGX Configuration Interface instruction sets by access method 26 Figure 3 3 Wi Fi access method for RGX Configuration Interface 28 Figure 3 4 Ethernet access method for RGX Configuration Interface 31 Table 3 5 Internal indicator lights and their meanings 31 Figure 3 5 iNet settings 32 Table 3 6 Power button light...

Page 6: ...re A 1 B Control drawing 1810D9509 200 revision 1 54 Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 7: ... des emplacements dangereux uniquement en ce qui concerne la sécurité intrinsèque par schéma de commande 1810D9509 200 WARNING Potential electrostatic charging hazard Only clean using a damp cloth AVERTISSEMENT Risque potentiel de charge électrostatique Nettoyez uniquement avec un chiffon humide WARNING Substitution of components may impair intrinsic safety and may cause an unsafe condition AVERTI...

Page 8: ...i même si le brouillage est susceptible d en compromettre le fonctionnement The RGX Gateway s optional Magnet Mount Kit and individual kit items can cause injury To avoid injury Industrial Scientific recommends the following x Persons with a pacemaker or implantable cardio defibrillator ICD should maintain a minimum separation distance of 90 cm 36 between the pacemaker or ICD and the magnet Please...

Page 9: ...GX U90 SM220 SQGBL652 Z64 CC3102MOD and RI7LE910SV USA ISED Canada 20727 RGX 7084A SM220 3147A BL652 451l CC3120MOD 5131A LE910NA and 5131A LE910SV Canada Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gayga...

Page 10: ...Y APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 11: ...s readings to gas alarms man down events panic alarms and more This wirelessly connected live monitoring system as depicted below enhances the safety team s quick responses to and preparedness for hazardous events iNet Instrument data RGX Gateway iNet Now users live monitoring and real time alerts Figure 1 1 System overview Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPR...

Page 12: ... more than one option is on the customer prioritizes the order in which the unit will use them For example if Wi Fi and cellular are on the unit can be set to first use Wi Fi and if that is not available use cellular The RGX iNet data exchanges are automatically encrypted using the Industrial Scientific encryption key Optionally a customer supplied key may be used Mobility and mounting The versati...

Page 13: ...r supplied conduit hub fixtures as needed for the connection of a compatible fixed DC power source Ethernet cable or both Compatibilities Instruments Use the information supplied below to ensure gas detection instruments are compatible with the RGX Gateway As needed upgrade instrument firmware to the required version add LENS Wireless functionality and refer to the instrument s product manual Tabl...

Page 14: ...put power parameters are provided below Table 1 3 RGX Gateway intrinsic safety IS input parameters Item Value Ui 16 2 VDC Ii 0 980 A Pi 2 2 W Ci 0 μF Li 0 mH The RGX Gateway can be operated using customer supplied customer installed fixed DC power that supplies a compatible input voltage range of 9 30 VDC with a maximum current of 5A and is connected to the dedicated ports on the RGX terminal bloc...

Page 15: ...ory setting values Status indicators Four x COMM indicates communication status x LENS indicates LENS Wireless status x ERR2 indicates communication error x ERR1 indicates system error Ethernet indicators Two x Yellow indicates connection status x Green indicates connection speed Operating conditions Ambient temperature 20 C to 55 C 4 F to 131 F Humidity 5 95 relative humidity RH noncondensing con...

Page 16: ... F and above 45 C 113 F Power supply accessory run time effects The RGX Gateway can draw operational power from compatible power supply accessories only from Industrial Scientific Provided below are the run time estimates for the RGX when it is used with each accessory Each power supply accessory has its own use restrictions Refer to an accessory s Product Manual to determine if the power supply s...

Page 17: ...ng power supply port and tethered cap Vent Tether chain screw Lock tabs for use with customer supplied item Activation code Top Bottom Handle Conduit hub plug 2x each covers a hole size of 27 8 mm 1 094 Figure 1 2 A Hardware overview exterior Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CO...

Page 18: ...mpatible input voltage range of 9 30 VDC with a maximum current of 5A and is connected to the dedicated ports on the RGX terminal block bWhen the factory reset button is pressed and held for approximately 20 seconds the unit s settings will revert to the original factory setting values After a reset adjust unit settings as needed Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENC...

Page 19: ...hould be decided for deployment planning and setup preparation x Decide which named LENS group e g Group A the RGX will monitor x A LENS group can include up to 25 equipment items counting gas detection instruments and RGX units x More than one RGX unit can be set to monitor a single LENS Group iNet manages for any duplication of instrument data sent from multiple RGX units RGX Gateway iNet commun...

Page 20: ...om Menu Wireless Setting LENS Wireless Encryption Valuec Automatic or custom Menu Wireless Setting Encryption Valuec Default or custom aAll equipment items must be set to the same LENS Wireless group and encryption key b X is used here to indicate the value will be a single character in length the available values are the letters A through J c Default sets the unit s LENS Wireless to the Industria...

Page 21: ... positioned to face the other instrument To learn more about the Ventis Pro or the Radius BZ1 consult its product manual Sample Deployment Plans As illustrated below in simplified two dimensional sample deployment plans LENS provides flexibility for iNet Now live monitoring to occur when fully enabled wirelessly connected instruments are located within a concentrated area spread out over a distanc...

Page 22: ...e 2 1 A LENS Wireless Group A instruments monitored all Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 23: ...1 B LENS Wireless Group A instruments monitored all but one Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 24: ...e 2 1 C LENS Wireless Group B instruments monitored all Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 25: ...1 D LENS Wireless Group B instruments monitored all but one Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 26: ...Y APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 27: ...s to have been damaged contact Industrial Scientific or an authorized distributor of Industrial Scientific products After unpacking peel the protective plastic liner from the RGX case lid Table 3 1 Package contents Quantity Item Description As ordered RGX Gateway Moves data between compatible enabled Industrial Scientific gas detection instruments and iNet As ordered Charging power supply and powe...

Page 28: ...RGX Serial Number S N which is located on the right side of the unit x RGX Activation Code A C which is located on the right side of the unit Log into iNet and follow this navigation x Click on the Equipment tab x Choose Activate New Equipment x In the provided fields enter the unit s Serial Number and Activation Code x Use the OK button to enter the information Gas detection instruments For gas d...

Page 29: ...power cord Connect the charging power supply to its power cord Uncap the unit s charging power supply port turn the port cap counterclockwise to expose the port for use Connect the power supply cable to the port and turn its swivel cover clockwise approximately 45 until it clicks closed or a full stop connection feedback is felt Plug the power cord into a compatible outlet Document Number 00000375...

Page 30: ... not blinking green x For information about other light patterns and colors see Table 3 6 Power button light indicators Shutdown Press the power button for approximately 7 seconds x When the power button light starts to blink release the button x To indicate unit shutdown is in progress the light will blink red and green x To indicate the unit is fully shutdown the light will turn off Configuratio...

Page 31: ...iguration Interface the following apply x If you are not familiar with your company s communication operations see your network administrator for assistance x If your company s wireless security network is WPA2 Enterprise the RGX Interface will require the upload of a certificate of authenticity that is 2048 bit or less and supplies the full chain of trust x As with most applications the Interface...

Page 32: ...et No Yes Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 33: ...iguration Interface x User name RGXadmin case sensitive x Password Unit activation code Note It is recommended that the password be changed after initial log in 4 Configure communication options Access Settings Review the RGX iNet communications options cellular Wi Fi and Ethernet x Enable or disable each option x For each enabled option enter select or edit the required values Reminder If your co...

Page 34: ...nstructions to use iNet where you will complete all other RGX settings Figure 3 3 Wi Fi access method for RGX Configuration Interface Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 ...

Page 35: ...GX default static IP is 192 168 1 1 Next connect one end of the customer supplied Ethernet cable Cat5 or greater to the computer then connect the other end to the RGX Ethernet port Power on the RGX On the case lid press the power button for approximately 5 seconds Observe the power button light and proceed as noted below Green only blinking or solid Continue Includes red blinking or solid See Tabl...

Page 36: ...ed click on save you will then be prompted to confirm the save When prompted to reboot simply click on the request then confirmation You do not need to do anything to the unit as the software will complete the reboot 7 Status check Observe the board s LEDs inside the case and compare them to the operational status column in Table 3 5 below x The LENS Wireless status light will be off until another...

Page 37: ...ss status Green LENS Wireless is operational if blinking green data are being exchanged Off Either LENS Wireless is not operational or no other equipment items are in the LENS group ERR2 Communication status Red There is a communications error Off There is no communications error ERR1 System status Red There is a system error Off There is no system error Ethernet cable port Left LED Ethernet conne...

Page 38: ... sites 3 If the gas detection instruments have not yet been activated for live monitoring follow this navigation x Click on the iNet Now tab x Choose iNet Now Activations Use the serial number search box to locate an instrument A check mark in the Activated column indicates an instrument is activated for live monitoring Figure 3 5 iNet settings Installation Precheck Power on the unit if it is not ...

Page 39: ...and the unit is shutting down Only red is on Solid not blinking The unit is not operational because there is no power or there is a system fault Light is off The unit is powered off or the battery is completely discharged Status is indicated only when feature is enabled Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In W...

Page 40: ...Y APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 41: ...ws these examples x NFPA 70 National Electric Code NEC Article 504 x CSA C22 1 Canadian Electric Code CEC Appendix J 18 152 x IEC EN 60079 14 Electrical installations A list of possible customer supplied equipment items and services is provided below Table 4 1 Customer supplied equipment and services Item Purpose Restrictionsa Mounting optional Mounting surface and mounting equipment items Mountin...

Page 42: ... suitable for extended run time installations Before using the ERTPS ensure it suits your application read and understand its Product Manual part number 17158385 Note Except for its IS cable the ERTPS is not certified for use in any hazardous classified areas Supplies optional Torx screwdriver set with T20 T25 and T30 bits For use with removal and installation of conduit hub plugs T20 screwdriver ...

Page 43: ...m distance of 5 m 16 4 between the RGX and the accessory or any nearby ISERTPS or ERTPS accessories see Figure 4 1 below This allows the RGX to be operated in a manner designed to prevent radio frequency interference per FCC Federal Communications Commission requirements x If the RGX will be connected to a compatible customer supplied customer installed fixed DC power source Ethernet cable or both...

Page 44: ...xed DC power source Installation Preparation Based on how the RGX will be mounted or placed and its intended operational power source and communications options use the information below to determine which preparation tasks are needed Refer to the Service section of this Product Manual for task instruction Table 4 2 Possible installation preparation tasks Application Service task Mounting Wall sur...

Page 45: ...permitted to be connected at a given time 12V Charger Port IS Power Port or 9 30 VDC Terminal Block Based on the operational power the unit will use and whether or not an Ethernet connection will be installed complete only one of two possible instruction sets as indicated below in Table 4 3 Read the full instruction set before starting Along with the information supplied in the instruction set mee...

Page 46: ...arger Port IS Power Port or 9 30 VDC Terminal Block If the RGX will draw operational power from a compatible power supply accessory cap the RGX charging power supply port Ensure the ISERTPS or ERTPS is installed connected and used in accordance with the instruction provided in its product manual Skip to Operation Precheck c Charging power supply part number 17158665 use only in a nonhazardous loca...

Page 47: ...installing a port cap push the cap slightly then turn its swivel connector clockwise approximately 45 until it clicks closed or a full stop connection feedback is felt 4 Label check The Wall Mount Kit includes a label that replicates important information that appears on the back of the RGX which includes the unit s LAN MAC and WAN MAC values These two values are to be customer recorded onto the k...

Page 48: ...choose only one instruction set below a or b but not both to connect only one external power source to the RGX WARNING Only one of the following power input ports is permitted to be connected at a given time 12V Charger Port IS Power Port or 9 30 VDC Terminal Block a Power supply accessory ISERTPS or ERTPS WARNING Only one of the following power input ports is permitted to be connected at a given ...

Page 49: ...ide by its tab Place the cable or power line in the guide s groove Lower and secure the tab to the guide base 10 Close case lid and secure to case base Ensure all wires and cables are secure and are contained inside the case and away from the unit s closure surfaces This will support the integrity of the items and full closure of the unit If the strain relief on the Ethernet cable resists full clo...

Page 50: ...t patterns refer to Table 3 6 Power button light indicators for more information Document Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1...

Page 51: ... protective gloves and eyewear to avoid a potentially severe pinch injury cut or splinter Service tasks that can be completed by Industrial Scientific customers are described in this Product Manual Table 5 1 indicates which parts and components are customer replaceable All other service tasks should be performed only by Industrial Scientific or its authorized service centers Supplies Torx screwdri...

Page 52: ...ower port cap 18109571 Includes tether chain screw sealing gasket washers and hex nut 4 Tether chain screw Screw torque 0 63 Nm 90 oz in 5 Vent 18109572 Vent nut torque 1 27 Nm 180 oz in 6 Conduit hub plug 18109574 Screw torque 1 27 Nm 180 oz in Accessories Wall Mount Kit 17157620 Screw torque 1 13 to 1 69 Nm 160 to 240 oz in Magnet Mount Kit 18109564 Do not use with a unit that will draw power fr...

Page 53: ...ver turn the torx screw clockwise to tighten Screw torque 1 13 to 1 69 Nm 160 to 240 oz in Install the remaining wall mounts In the kit locate the supplied label which replicates important information that appears on the back of the RGX Locate the unit s LAN MAC and WAN MAC values which are included on the label that is factory affixed to the back of the unit Record these two values onto the kit l...

Page 54: ... the port for use Align the wide tab of the metal adapter connector to the wide notch on the IS port push on to the port to connect Turn the swivel connector clockwise approximately 45 until it clicks closed or a full stop connection impact is felt Connect the other end of the adapter in accordance with the Product Manual instruction from the compatible Industrial Scientific power supply accessory...

Page 55: ... bottom of the case base Inside the case hold the nut in place with a 3 8 open end wrench and turn the T20 torx screw counterclockwise to loosen Remove the nut two internal washers the sealing gasket and one external washer from the case To remove a worn or damaged fixed gasket push the outer edge of the gasket toward its center with a finger use friction and a forward motion to separate the gaske...

Page 56: ...base From outside of the case place the torx screw into the concave side of the washer and place over the base opening Inside the case place the next washer convex side down over the screw and fixed gasket Follow with the sealing gasket flat washer and hex nut Hold the hex nut in place with a 3 8 open end wrench and turn the T20 torx screw clockwise to tighten Screw torque 1 27 Nm 180 oz in Docume...

Page 57: ...ems according to company policy Port cap installation Hex nut Flat washer Copper ground shield Flat washer Sealing gasket Connect the replacement port cap to the appropriate port IS port or Charging port turn clockwise approximately 45 until it clicks closed or a full stop connection impact is felt Insert the T10 torx screw through both tether chain end rings and into the case screw hole Always ca...

Page 58: ...ISED OF THE POTENTIAL FOR SUCH DAMAGES Industrial Scientific s total liability hereunder from any cause whatsoever except liability from personal injury caused by Industrial Scientific s negligence whether arising under contract warranty tort including negligence strict liability products liability or any other theory of liability will be limited to the lesser of Buyer s actual damages or the pric...

Page 59: ...ent Number 0000037533 DO NOT CHANGE WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 60: ... WITHOUT THE FOLLOWING AGENCY APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 61: ...Y APPROVALS S N A Modified 2018 05 25 13 18 14 EDT State In Work INDUSTRIAL SCIENTIFIC CORPORATION Title PRODUCT MANUAL RGX GATEWAY EN 1 Eng Drawing No 17158071 1 Drawn By Gaygan Nancy Created 2018 05 16 10 24 51 EDT Rev 1 Printed By Kuzmich Bob Int 1 5 Date May 29 2018 ...

Page 62: ...otre site Web www indsci com si vous voulez trouver un distributeur de nos produits près de chez vous ou si vous recherchez un centre de service ou un bureau Industrial Scientific Besuchen Sie uns unter www indsci com um einen Vertriebshändler unserer Produkte oder ein Servicecenter bzw eine Niederlassung von Industrial Scientific zu finden Para buscar un distribuidor local de nuestros productos o...

Reviews: