Basic Operation of the DS2
DS2 Docking Station
206
INDUSTRIAL SCIENTIFIC
Version 9.1 (P/N: 17112798)
Keep IDSs away from radiators and heat sources. Do not block cooling vents.
Avoid
placing loose papers underneath an IDS; do not place an IDS in a closed-in wall unit, or
on a bed, sofa, or rug.
An IDS is equipped with a fixed-voltage power supply. The IDS will operate at only one
voltage (see the regulatory label on the outside of the IDS for its operating voltage).
7.8.2. Cleaning
Before you clean your IDS, disconnect the power cord from the electrical outlet. Clean your IDS
with a soft cloth dampened with water. Do not use liquid or aerosol cleaners, which may contain
flammable substances. Do not spray water directly onto the unit.
7.8.3 Explanation of Symbols Used on Unit
Symbol
Description
Direct Current
Alternating Current
Protective Conductor Terminal
Caution (refer to accompanying documents)
7.8.4 Specifications
CAUTION: Equipment is rated for indoor use only. Use only indoors at altitudes below 2000
meters (6000 ft.)
Electrical Input:
+12VDC +/-500mV, 5A
100-240 VAC 50/60Hz, 1.75A (optional)
Note:
Mains supply voltage not to exceed ±10% of nominal voltage.
Installation Category #2
Operating Temperature Range:
5°C to 40°C
Operating Humidity Range:
0 - 80%RH up to 30°C, decreasing linearly to 50%RH at
40°C
Pollution Degree 2
7.8.5. Regulatory
Notices
Electromagnetic Interference (EMI) is any signal or emission, radiated in free space or conducted
along power or signal leads, that endangers the functioning of radio navigation or other safety