Default Settings
DS2 Docking Station
232
INDUSTRIAL SCIENTIFIC
Version 9.1 (P/N: 17112798)
11.4. Default Calibration Gases
DS2 Docking Station allows you to define the concentration of calibration gas that you typically
use to calibrate certain sensor types. You can define calibration gases for each of the Docking
Station compatible instruments.
Default Calibration Gases apply only to instruments when they are docked for the first time, or to
new sensors whose information is not in the database. Therefore, you should set up your default
settings before docking any instruments that are registered in the database. Changing the Default
Calibration Gas settings after an instrument has been registered will not update the instrument
sensor’s settings, unless a new sensor is installed in the instrument.
11.5. Modifying Default Calibration Gas Settings
To modify Default Calibration Gas settings, follow the instructions listed below.
Step Instruction
1.
Open the DSSAC application.
2.
Click on the View menu and select Default Calibration Gases.
3.
The Company Calibration Gases dialog box appears. The Company Calibration
Gases dialog box contains a tab for each Docking Station compatible instrument.
Each tab contains a list of sensor types that applies to the instrument. Select the
instrument for which you want to change calibration gas settings.
Figure 11-3. Company Calibration Gases Dialog Box
NOTE:
VX500 Photo Ionization Detector contains one entry for the PID Sensor.