DS2 Docking Station
Configuring the DS2
Version 9.1 (P/N: 17112798)
INDUSTRIAL SCIENTIFIC
193
Figure 6-16. Diagram of iGas Smart Card and Reader
NOTE:
It is important that the gas line is connected before the iGas card so that the system is
ready to draw gas after it reads the card.
NOTE:
Be sure that the Smart Card is connected to the correct iGas Port. For example, if you
have connected the gas line to Gas In # 2, then the Smart Card must be connected to iGas Port #2.
If the correct port is not used, the system could use the wrong type of gas for a calibration or
bump test, rendering the results inaccurate.
Step Instruction
9.
The system reads the information about the gas cylinder from the iGas Smart Card
and automatically configures the Gas In connection in the DSSAC.
10.
The gas cylinder is ready to use.
NOTE:
If you are using iGas, do not edit the cylinder configuration in DSSAC. Doing so could
temporarily override the settings configured by iGas, possibly causing DS2 Docking Station to
use inaccurate information for calibrations or bump tests. If the
settings do get overwritten,
simply reinsert the iGas Smart Card to reconfigure the gas cylinder.