Reset SF
6
Density Trend Data
If an SF
6
density sensor is found to be faulty, the SF
6
Trend
calculation should be re-started after replacing the sensor.
This can be done by resetting the accumulated SF
6
data
for that channel. When this action is taken, a new array
of data will be accumulated and a new SF
6
Trend will be
calculated for that channel. Until there is a full array of data
to calculate the new Trend, the Confidence Level for that
channel will indicate less than 100%.
To reset the SF
6
Trend Data
for a channel
, select the
ACTION tab and click “
Reset
SF
6
Trend Data and Alarm
Channel A, B or C
”. A yellow bar will appear at the top of
the window.
Click “YES” to confirm the action. The correct
Administrator’s password will need to be given for the
change to be saved. Click “Apply” to complete the action.