3.9.3 Download .DXP File
To download the .dxp file via Ethernet, select the menu item
Download
in your browser.
Please note:
This menu item is only displayed in your browser and not in the RCI.
Click on the file to be downloaded from the file list to import it. (Default:
DEFAULT.dxp)
If there is no network connection to the computer, remove the SD card of the D
X
and insert it
into your computer’s card reader. The .dxp files are saved in the root directory.
3.9.4 Automatically load Configuration File when booting
When booting, the D
X
automatically imports the configuration file DEFAULT.dxp. If you save
your current configuration under this file name, it is automatically imported when rebooting.
Select
Load/Save
and type in
DEFAULT
as File Name.
Select the
Save
button.
Confirm the message.
The configuration selected is now automatically loaded at restart.
3.10 Start Measurement
3.10.1 Measurement Mode and Configuration Mode
When the D
X
RCI is switched on, it is always in one of the following modes:
In measurement mode, data is recorded and transmitted.
In configuration mode, the SCTs and the RCI can be programmed. Here, no
measurement data is transmitted.
To ensure consistent data, you immediately exit measurement mode when leaving the menu
page
Measure → Start
(input by scroll wheel or Ethernet)
Please note:
If you call the D
X
configuration menu in a new browser session, you are
automatically redirected to the D
X
main page. The system then interrupts data transmission
and switches to configuration mode.
To prevent any access to the D
X
Ethernet interface during measurements, you may unplug
the Ethernet cable.