D
x
Telemetry System
Manual Vers. 2.7
83
/
92
9
Maintenance and Storage
9.1
Cleaning and Maintenance
Due to its construction, the telemetry system is protected against contamination of its
electronic parts. Contaminations of the case can be removed with non-agressive agents like
isopropyl alcohol when the device is switched off.
Due to constructional design of D
X
transmitter and receiver, there is no maintenance
required, if cables are protected from mechanical stress.
9.2
Sensor Input Calibration
CAEMAX Technologie GmbH recommends yearly calibration of SCT sensor inputs. For
further information, please contact
[email protected]
9.3
Storage
The telemetry system can be stored for later usage. However, it has to be protected from:
Electrostatic charging (may destruct electronic parts)
Humidity (leads to corrosion)
Permissible storage temperature: -10...40
C
Avoid tangles when storing cables. The modules must not be damaged mechanically when
being packed. Batteries have to be stored separately from the modules, so that the latter
cannot be damaged in case of leaking batteries.
9.4
Customer Cable Modification
When lengthening and/or altering cables and plugs on your own, keep in mind that you have
to provide sufficient insulation and to connect adequately. In case the customer carries out
cable modifications on his own, the manufacturer’s function guarantee expires.