5. Setting up the device
14
5.4. Extended parameterisation
(Professional operation level)
Menu level
Parameterisation level
Selection of the input signal,
tYPE
:
Default:
fre9u
If the scaling of the device is done via
SEnS.F
(Sensor calibration), the frequency range needs to
be preset under
rAnGE
and is adjusted by application of the final value/initial value. If
FrEqU
(factory calibration) is preferred, the final value needs to be entered under
End
and the final
frequency needs to be entered under
EndA
. Under
OFFS
the initial value needs to be entered and
under
OFFSA
the initial frequency. There is no application of the measuring signal. Confirm the
selection with
[P]
and the display switches back to menu level.
Setting the end value of the measuring range,
END
:
Default:
100e3
Choose between six different frequency ranges. Confirm the selection with
[P]
and the display
switches back to menu level.
Setting the final value of the measuring range,
END
:
Default:
10000
Set the final value from the smallest to the highest digit with
[
▲
] [
▼
]
and confirm each digit
with
[P]
. A minus sign can only be parameterized on the highest value digit. After the last digit,
the display switches back to the menu level. If
Sens
was selected as input option, you can only
select between
noca
and
cal.
With
noca
, only the previously set display value is taken over, and
with
cal
, the device takes over both the display value and the analogue input value.
Menu group level
Menu level
5.4.1. Signal input parameters
9.9999 Hz
99.999 Hz
999.99 Hz
9.9999 kHz
99.999 kHz
999.99 kHz