5. Setting up the device
Menu level
Parameterisation level
Function if display falls below / exceeds limit value,
FU-1:
Default:
high
The limit value undercut can be selected with
Louu
(LOW = lower limit value) and limit value
exceedance can be selected with
high
(HIGH = upper limit value). If e.g. limit value 1 is on a
switching threshold of 100 and occupied with function „
high
“, the alarm will be activated by
reaching the threshold. If the limit value is allocated to „
Low
“, an alarm will be activated by
undercut of the threshold.
See page 29
.
Threshold values / limit values,
LI-2:
Default:
3000
For both limit values, two different values can be parameterized. With this, the parameters for
each limit value are called up one after another.
Hysteresis for limit values,
HY-2:
Default:
00000
The delayed reaction of the alarm is the difference to the threshold value, which is defined by
the hysteresis.
Function if display falls below / exceeds limit value,
FU-2:
Default:
high
The limit value undercut can be selected with
Louu
(LOW = lower limit value) and limit value
exceedance can be selected with
high
(HIGH = upper limit value). If e.g. limit value 1 is on a
switching threshold of 100 and occupied with function
high
, the alarm will be activated by
reaching the threshold. If the limit value is allocated to
Low
, an alarm will be activated by
undercut of the threshold.
User code (4-digit number-combination, free available),
U.CodE
:
Default:
0000
If this code is set (>0000), all parameters are locked, if
LOC
has been selected before under
menu item
run
. By pushing
[P]
during operation mode for approx. 3 seconds,
Code
appears in
the display. To get to the unlocked reduced parameter, the user needs to enter the preset
U.Code
. This code has to be entered before each parameterisation, until the
A.Code
(master code)
unlocks all parameters again.