3-28
3
TABLE 3-3 SCPI COMMANDS AND QUERIES
(Continued)
Keyword
Default
Description
Value
ber and firmware revision) separated by
commas. e.g. ICS Electronics, 4807, S/N
012345, Rev 1.1 (04-08-00).
:DATe <date>
Saves IDN message and date. The save
operation lights all the LEDs. Date is in
mm/dd/yy format.
:DATe?
Queries the calibration date. The response
is 00/00/0000 for factory default settings.
DEFault
Sets Flash memory to factory settings.
Does not affect A/D. Temperature or D/A
calibration values.
:LOCK
0
Disables configuration commands when
On. Values = 0|1 or OFF|ON. Table 1-4
lists the locked commands.
:MEASure
Calibrates A/D input channels.
:OFFset
0
Sets A/D zero offset for the current se-
lected range. Parameter is Channel Num-
ber, Value. Value is 0 to 1000. e.g.
CAL:MEAS:OFFset 2, 0.004
:GAIN
1
Sets positive input A/D scale factor for
the current selected range. Command
format is Channel number, value. Value
is 0 to 1000. e.g. CAL:MEAS:GAIN 2,
1.01
:NGAIN
1
Sets negative input A/D scale factor for
the current selected range. Applies to
bipolar ranges only. Command format is
Channel number, value. Value is 0 to
1000. e.g. CAL:MEAS:NGAIN 2, 1.01