TABLE 3-3 SCPI COMMANDS AND QUERIES
(Continued)
Keyword
Default
Description
Value
LISTen
Sets listen string and data output format.
ASCii converts a decimal value into an
eight bit binary bit pattern by bit weights.
Multiple byte values are separated by
commas. HEX converts the incoming
ASCII characters 0-9 and A-F into the
equivalent four bit HEX values. HEXL
allows the user to insert commas between
hex byte values. Values are ASCii | HEX
| HEXL
i.e.
ASCii example = 161,35,69
HEX example = A12345
HEXL example = A1,23,45
SOURce
Starts Voltage/Current Source branch
:VOLTage
Selects/sets voltage output branch for each
analog output. When a value is included
with the command, it is converted to an
output value by the Vout=M*Vcmd+B
equation. Final value must not exceed the
programmed limit value or the maximum
value for the selected range. When a
Bipolar range is selected, the value is
from the -maximum to the +maximum
limit. Limit checking is done on an abso-
lute value basis. Command format is
Channel number, value. Channel number
is 1 to 4. Value = 0 to 1000.
[:LEVel]
Optional voltage amplitude word
[:IMMeditate]
Optional voltage amplitude word
[:AMPLitude]
Optional voltage amplitude word. Com-
mand format is Channel number, value.
Channel number is 1 to 4. Value = 0 to
1000.
:TRIGgered
Selects/sets a delayed voltage output