![Hydronix Hydro-Control VI Operator'S Manual Download Page 26](http://html.mh-extra.com/html/hydronix/hydro-control-vi/hydro-control-vi_operators-manual_2192336026.webp)
26 Hydro-Control VI Operators Guide HD0456 Rev 1.7.1
3 AUTO Mode
The AUTO Mode progressively adds water to the mixer to achieve a defined moisture target. The
Hydro-Control controls the speed at which the water is added and slows the addition as the current
moisture nears the target moisture in order to accurately reach the target without overshooting it.
Moisture
Time
Mixer
Loading
Dry Mix
Time
Main Water Addition
Wet Mix Time
Discharge
Target
Current moisture
monitored and water
added to reach the target
moisture
Figure 10: The moisture during AUTO Mode
Figure 10 shows a typical moisture trace during a batch run in AUTO Mode.
As there is no requirement for complete homogeneity during the Dry Mix Phase, the Dry Mix Time
can be made shorter than in CALC Mode so the water can start to be added earlier in the batch.
The system is shipped with default parameters for controlling the water addition. To optimise the
system these settings may require adjustment.
To ensure that the moisture is correct and that the mix is homogeneous, it is important to allow a
sufficient Wet Mix Time before discharging the mixer.
4 Selecting the Best Mode – AUTO or CALC?
The most appropriate mode for moisture control varies depending on the application. It is important
to understand the differences between AUTO Mode and CALC Mode in order to select the mode
that is the most appropriate.
•
Do the batch sizes vary from one batch to another?
If yes then AUTO Mode will work without needing to have an exact dry weight parameter
in the recipe before a batch is made. To use CALC Mode the dry weight recipe parameter
would have to be updated on a batch by batch basis, either by sending it using the RS232
port on the Hydro-Control VI, or by manually editing the recipe.
•
Is the water supply pressure constant?
The AUTO Mode control method requires water to be added consistently. If there are
changes in flow rates due to pressure fluctuation, then the AUTO Mode will not be able to
perform optimally.
•
Are mixing times critical?
If yes then it is likely that CALC Mode may be quicker to use than AUTO Mode.
•
Is it possible to achieve a stable reading in the dry mix time?
A prerequisite of CALC Mode is that the water calculation is based on a correct and stable
dry mix reading for addition in a single dose. If the dry mix reading is not stable enough
then it is unlikely that CALC Mode will give the desired accuracy. AUTO Mode does not
need a stable signal due to water being added continuously to reach the target.
Summary of Contents for Hydro-Control VI
Page 4: ...4 Hydro Control VI Operators Guide HD0456 Rev 1 7 1 ...
Page 6: ...6 Hydro Control VI Operators Guide HD0456 Rev 1 7 1 ...
Page 8: ...8 Hydro Control VI Operators Guide HD0456 Rev 1 7 1 ...
Page 10: ...10 Hydro Control VI Operators Guide HD0456 Rev 1 7 1 ...
Page 18: ...Navigation Chapter 2 18 Hydro Control VI Operators Guide HD0456 Rev 1 7 1 ...
Page 20: ...Plant Operator Controls Chapter 3 20 Hydro Control VI Operators Guide HD0456 Rev 1 7 1 ...
Page 34: ...Editing Recipes Chapter 6 34 Hydro Control VI Operators Guide HD0456 Rev 1 7 1 ...
Page 40: ...Running the First Mix Chapter 7 40 Hydro Control VI Operators Guide HD0456 Rev 1 7 1 ...
Page 56: ...Using Moisture Control Chapter 8 56 Hydro Control VI Operators Guide HD0456 Rev 1 7 1 ...
Page 70: ...Setting Up User Accounts Chapter 11 70 Hydro Control VI Operators Guide HD0456 Rev 1 7 1 ...
Page 72: ...Remote Support Chapter 12 72 Hydro Control VI Operators Guide HD0456 Rev 1 7 1 ...
Page 86: ...Document Cross Reference Appendix C 86 Hydro Control VI Operators Guide HD0456 Rev 1 7 1 ...