GSC53N - GSC57 - ZG47
EN - 17
6.1.2. Measurement Procedure
1. Select the desired mode by means of the
F1
key.
2. Connect the black and blue test leads to
B1
and
B4
input terminals respectively
B2
B3
B4
B1
Connection of the test leads during LOW test.
3. If the cables supplied with the instrument are not long enough for the measurement
you can extend the blue cable.
4. Short-circuit the test leads making sure that the conductive parts of the crocodiles
make a good contact to each other. Press the
START
key.
If the display doesn't
show 0.00 repeat the test leads calibration
(see paragraph 6.1.1).
5. Connect the instrument terminals to the ends of the conductor under test (see
previous picture).
6.
If the mode "RT+" or "RT-"
was selected use the
F3
,
F4
keys to set the
duration of the test.
7. Press the
START
key. The instrument will execute the measurement. In RT+/RT-
(Timer mode) you can press
START
key again if you want to stop the test before the
duration set is expired.
WARNING
Never disconnect the test leads when the message
"MEASURING"
is
displayed.
START
STOP
Summary of Contents for GSC53N
Page 1: ...Copyright HT ITALIA 2012 Release EN 1 08 28 02 2012 GSC53N GSC57 ZG47 Instruction Manual...
Page 2: ......