GSC53N - GSC57 - ZG47
EN - 52
If a contact voltage
U
t
higher
than
the
selected limit (U
L
)
is
detected the instrument
interrupts the test and
emits
a long sound
signal
at the end of the
test and displays the
screen alongside.
LOOP
05.06.01
- - -
---A
FRQ=50.0HZ
VP-N= 1V Vp-PE= 0V
Ut DANGEROUS
R
A
50V
Message
“Ut”
:
the
instrument doesn't detect
an
efficient
protection
circuit.
FUNC
UL
If the instrument detects
such an high earth
resistance that the earth
conductor or the earth
installation itself seem to
be
unrealiable,
the
message alongside is
displayed. Check the
efficiency
of
the
protective conductor as
well
as
the
earth
installation.
LOOP
05.06.01
- - -
---A
FRQ=50.0HZ
VP-N=231V Vp-PE= 40V
NO PE
P-PE 50V
Message
“NO PE”
: the
instrument doesn't detect
an
efficient
protection
circuit.
FUNC Z2
UL
If the instrument gets
overheated, tests cannot
be carried out and the
message alongside is
displayed. Wait until the
initial screen is back in
order to proceed with
measurements.
LOOP
05.06.01
- - -
---A
FRQ=50.0HZ
VP-N=231V Vp-PE= 40V
HOT
P-P
Message
“hot”:
the
instrument got overheated.
FUNC Z2