GSC53N - GSC57 - ZG47
EN - 92
8.6.
"POWER" FUNCTION
This function permits you to display in real time the RMS value of AC/DC voltage, the peak
and ThdV value of the 3 phase voltages, the waveform of the 3 phase voltages, the RMS
value of AC/DC currents, the peak and Thdl of the 3 phase currents, the waveform of the 3
phase currents. Furthermore, the instrument calculates and displays the value of the
phase and total active powers, the value of the phase and total reactive and apparent
powers, the value of the phase and total power factors and cos .
8.6.1. Symbols
The position POWER has two working modes:
METER
WAVE
For voltage and current harmonics see paragraphs 8.4.3 and 8.5.3 respectively.
These modes will be described in detail in the next paragraphs.
The symbols used are described below:
Symbol
Description
V1, V2, V3
RMS value of the voltage of phase 1, phase 2, phase 3 respectively
V12, V23, V32, V31
RMS Value of the phase to phase voltages
freq
Network frequency
Phseq
Phase sequence indicator
"123" correct
"132" inverted
"023" null voltage on the black wire
"103" null voltage on the red wire
"120" null voltage on the green wire
"100" null voltages on the red and green wires
"020" null voltages on the black and green wires
"003" null voltages on the black and red wires
I1, I2, I3
RMS value of the current of phase 1, phase 2, phase 3 respectively
IN
RMS value of the current of the neutral
Pt, P1, P2, P3
Values of the active power (total, of phase 1, phase 2, phase 3 respectively)
P12, P32
(only for 3 wires measurement) Value of the power measured by the Wattmeter 1-2
and 3-2 respectively (see paragraph REF _Ref530397951 \r \h 17.12.2).
Qt, Q1, Q2, Q3
Values of the reactive power (total, of phase 1, phase 2, phase 3 respectively)
Q12, Q32
(only for 3 wires measurement) Value of the power measured by the VAR meter
Va1-2 and 3-2 respectively (see paragraph REF _Ref530397951 \r \h 17.12.2).
St, S1, S2, S3
Values of the apparent power (total, of phase 1, phase 2, phase 3 respectively)
S12, S32
(only for 3 wires measurement) Value of the power measured by the VA meter Va1-
2 and 3-2 respectively (see paragraph REF _Ref530397951 \r \h 17.12.2).
Pft, pf1, pf2, pf3
Values of the power factors (total, of phase 1, phase 2, phase 3 respectively)
dPft, dpf1, dpf2, dpf3 Value of the cos (total, of phase 1, phase 2, phase 3 respectively)
Ead, Pd
Values of the Total Active Energy and Active Power On demand respectively
Table 9: symbols used in the position POWER
The symbols "
i
" and "
c
" stand for reactive powers (Q), power factors (Pf) and cos (dpf)
respectively inductive and capacitive.
Summary of Contents for GSC53N
Page 1: ...Copyright HT ITALIA 2012 Release EN 1 08 28 02 2012 GSC53N GSC57 ZG47 Instruction Manual...
Page 2: ......