6.6.3. "RMEAS" and "VDROP" faulty cases
if the instrument doesn't
notice the presence of
Power in the socket "Power
ONLY
FOR
10A",
it
visualizes
screens
represented to side. Check
if the C5700 cable is
connected to instrument
and if the power outlet is
fed.
LOW 10A
05.06.01
- - -
----A
NO POWER SUPPLY
RMEAS RLIM=0.150
FUNC
Working mode
If the instrument detects
the
External
Power
supply
adapter
connected to instrument
will show the message
displayed to side.
LOW 10A
05.06.01
- - -
----A
REMOVE POWER
RMEAS RLIM=0.150
Disconnect the External
Power Supply Adapter
FUNC
THE PREVIOUS RESULTS CANNOT BE SAVED
.
In the RMIS mode If the
value of the resistance
is higher than the limit,
the instrument emits an
acoustic
signal
and
displays a screen similar
to displayed to side.
LOW
05.06.01
Resistance value higher
than limit
0.323
9.12A
RMEAS RLIM=0.150
Test current
FUNC
The displayed result can be stored pressing the
SAVE
key
twice
(according to
paragraph 9.1).