9. SAVING RESULTS
The
SAVE
button can be used to store the displayed results related to the rotary switch
position:
SAFETY TEST
and for
AUX
rotary switch position: pressing this key the instrument will
store the displayed result generating a corresponding record in the SAFETY TEST
MEMORY (see paragraph 11.1)
ANALYSER
rotary switch position: pressing this key the instrument will store the
displayed result generating a "Smp" record in the ANALYSER MEMORY (see
paragraph 11.2)
Please note that Saving results is different from recording.
9.1.
SAVING SAFETY TEST RESULTS
After a
SAFETY TEST
(function LOW , M , RCD, LOOP, EARTH) or during a real time
measurement in
AUX
position the user can press the
SAVE
button to store the displayed
result.
The
REMINDER
PLACE
parameter
isn't
related
to
Measurement Order
Number and can help
the user to remind the
place
where
he
performed
the
measurement.
05.06.01
SAVE IN LOC. 006
OK? (SAVE/ESC)
REMINDER PLACE:009
Memory Location
REMINDER PARAMETER
The following keys are available:
F3
,
F4
: to adjust the REMINDER PLACE.
SAVE
:
to store the test result associating to the actual REMINDER PLACE
ESC
:
to quit this mode without saving.
9.2.
SAVING DISPLAYED VALUES OF ANALYSER FUNCTION
During a Real Time measurement (in
ANALYSER
function) if the user press the
SAVE
button, a "Smp" record will be generate in the "ANALYSER MEMORY". This file contains
the Voltage and Current values present at instrument's input when the user pressed the
SAVE
key.
Downloading these values to a PC (using the management Software) the Power, Energy,
Harmonics, etc values can be calculated and displayed as well.