GSC53N - GSC57 - ZG47
EN - 70
7.2.
ENVIRONMENTAL PARAMETER AND LEAKAGE CURRENT: RECORDING
Before starting a recording we recommend You to check that
real time values
are correct.
To this purpose follow the measurement procedure described in paragraph 7.1.
in addition it's fundamental that Instrument settings correspond to the accessories is use.
For this we recommend you to check instrument's setting before execute an AUX
recording.
To this purpose please check the RECORDER CONFIG settings.
MENU
:
to enter in the
MENU
mode and change the instrument settings. It’s
not possible to enter the configuration MENU during a recording or an
energy measurement.
START/STOP
: to record the
selected parameters
according to the instrument’s
settings (see chapter 10).
7.2.1. AUX Basic setting: RECORDER CONFIG
Place the rotary switch in the
AUX
position, press the
MENU
key, using the
F1
/
F2
keys
select the
RECORDER CONFIG
item and press the
ENTER
Key.
WARNING
It’s not possible to enter the
MENU
during a recording or a Real Time
Energy measurement.
MENU GENERAL
SAFETY TEST MEMORY
ANALYSER MEMORY
RESET
ANALYSER CONFIG
RECORDER CONFIG
CONTRAST
DATE&TIME
LANGUAGE
This option allows you to check and eventually modify the recording parameters and the
selected parameters (up to a maximum of 3). The RECORDER CONFIG mode is divided
into 2 separate sub-pages:
1
st
page:
This page allows you to set the START/ STOP mode (AUTO or
MANUAL), the START and STOP time (if AUTO mode is selected) and
the Integration Period value. Press
ENTER
to confirm the settings and
pass to the following page.
Press
ESC
to leave the Menu without modifying the existing parameters.
2
nd
page:
This page allows the selection of the Input channels to be recorded.
Press
ENTER
to confirm the settings and leave the RECODER CONFIG
Menu. Press
ESC
to leave this page without modifying the existing
parameters.