TABLE 5-1. RECOMMENDED TEST EQUIPMENT
1. HP MODEL 740B DC VOLTAGE STANDARD
2. HP MODEL 3310A FUNCTION GENERATOR
3. HP MODEL 410C or 427A DC VOLTMETER
4. HP MODEL 202A LOW FREQUENCY FUNCTION
GENERATOR
5-53. INITIAL CHECKS
5-54. To perform initial performance checks:
a.
Set 115/230 selector switch to appropriate power
source.
b.
Connect power cable.
c.
Install pen and paper.
d.
Position POWER/SERVO toggle switch to ON-ON.
e.
Position CHART toggle switch to HOLD. Paper is
held down securely.
f.
Position PEN toggle switch to DOWN.
g.
Connect DC Standard to high and low (+ and -) Y-axis
inputs.
h.
Apply positive dc voltage. Pen moves upscale.
i.
Return pen to zero position.
j.
Disconnect DC Standard from Y inputs. Connect to X
inputs.
k.
Apply positive dc voltage. Pen moves across
recording area.
5-55. Y-AXIS ACCURACY AND LINEARITY
a.
Connect DC Standard to Y input terminals.
b.
Place pen at exactly zero on Y-axis and at 5 inches
(12. 5 cm) on X-axis.
c.
Apply 0.7 volt to Y input terminals.
d.
Pen should stop at 7 inches (18 cm) within ±0.014 in.
(0.34 mm).
e.
If pen does not stop within tolerance, adjust
calibration control (R-126).
f.
Reduce voltage applied in steps of 0.1V at each step.
g.
After each step, pen should stop at 1 in. (2 cm)
intervals within 0.014 in. (0.34 mm).
5-56. X-AXIS ACCURACY AND LINEARITY
a.
Connect DC Standard to X input terminals.
b.
Place pen at exactly zero on X-axis and at 3.5 in. (9
cm) on Y-axis.
c.
Apply 1 volt to X input terminals.
d.
Pen should stop at 10 inches (25 cm) within ±0.020
in. (0.5 mm).
e.
If pen does not stop within tolerance, adjust
calibration control (R226).
f.
Reduce voltage applied in steps of 0.1V at each step.
g.
After each step, pen should stop at 1 in. (2 cm)
intervals within 0.020 in. (0.5 mm).
5-57. Y-AXIS SLEWING SPEED
a.
Hook up recorder as shown in Figure 5-12.
FIGURE 5-12. Y-AXIS SLEWING SPEED TEST SETUP
b.
Set Function Generator to 0.5 Hz and triangular wave
output with peak-to-peak output of 1V.
c.
Position pen to bottom grid line.
d.
Apply 0.7 VDC to Y input at beginning of an X
excursion.
e.
Measurement is shown in Figure 5-13.
5-58. X-AXIS SLEWING SPEED
a.
Hook up recorder as shown in Figure 5-14.
b.
Set Function Generator to 0.715 Hz and a triangular
wave output with peak-to-peak output of 0.7V.
c.
Position pen to bottom grid line.
5-8
Summary of Contents for HP-7035B
Page 1: ...TECHNICAL MANUAL HEWLETT PACKARD MODEL HP 7035B ...
Page 9: ...FIGURE 1 4 DIMENSION DRAWING FIGURE 1 5 ACCESSORIES 1 3 ...
Page 12: ...FIGURE 1 7 TYPICAL FREQUENCY RESPONSE INPUT FILTER FIGURE 1 8 TYPICAL DYNAMIC RESPONSE 1 6 ...
Page 14: ...FIGURE 2 1 WING BRACKET INSTALLATION FIGURE 2 2 STACKED CONFIGURATION 2 2 ...
Page 16: ...FIGURE 3 1 FRONT PANEL CONTROLS 7035B AND 7035B 001 MODELS 3 2 ...
Page 20: ...FIGURE 4 1 BLOCK DIAGRAM FIGURE 4 2 DETAILED BLOCK DIAGRAM 4 2 ...
Page 48: ...TM 11 6625 2850 14 P FIGURE 6 1 EXPLODED VIEW CABINET 6 9 ...
Page 49: ...TM 11 6625 2850 14 P FIGURE 6 2 EXPLODED VIEW CARRIAGE ARM SHEET 1 OF 2 6 10 ...
Page 58: ...FIGURE 7 1 COMPONENT IDENTIFICATION FRAME 7 8 ...
Page 59: ...TM 11 6625 2850 14 P Figure 7 2 Component Identification Circuit Board 7 9 ...
Page 60: ...TM 11 6625 2850 14 P Figure 7 3 Schematic Model 7035B D 07035 92550 7 10 ...
Page 62: ...TM 11 6625 2850 14 P FIGURE 8 1 COMPONENT IDENTIFICATION CIRCUIT BOARD 8 2 ...