7-4
Options and Accessories
Analyzer Options Available
Option 085, High Power System (ES Only)
This option is designed to permit the measurement of high power devices. With an
external power amplifier, this configuration will allow up to 20 Watts (+43 dBm) of output
at the test ports. The maximum test port input power is 1 Watt (+30 dBm) CW, but
jumpers on the front panel allow the insertion of high power attenuators or isolators. This
allows test device output levels up to the power limits of the inserted components.
Additionally, there is an external reference input that allows the external amplifier's
frequency response and drift to be ratioed out, and there are internally controlled step
attenuators between the couplers and samplers to prevent overload. A network analyzer
with this option can be configured to operate as a normal instrument (with slightly
degraded output power level and accuracy) or as an instrument capable of making single
connection multiple measurements. Because of high output power, Option 085 is only
available with a mechanical transfer switch similar to Option 007. Option 085 also
includes direct access to the receiver (Option 012).
Option 089, Frequency Offset Mode (ES Only)
This option adds the ability to offset the source and receiver frequencies for frequency
translated measurements. This provides the instrument with mixer measurement
capability. It also provides a graphical setup that allows easy configuration of your mixer
measurement.
Option 400, Four-Sampler Test Set (ES Only)
This option reconfigures the instrument's test set to ratio out the characteristics of the test
port transfer switch, and to include a second reference channel that allows full accuracy
with a TRL measurement calibration.
Option 1CM, Rack Mount Flange Kit Without Handles
Option 1CM is a rack mount kit containing a pair of flanges and the necessary hardware to
mount the instrument, with handles detached, in an equipment rack with 482.6-mm (19
inches) horizontal spacing.
Option 1CP, Rack Mount Flange Kit With Handles
Option 1CP is a rack mount kit containing a pair of flanges and the necessary hardware to
mount the instrument with handles attached in an equipment rack with 482.6 mm (19
inches) spacing.
Service and Support Options
Hewlett-Packard offers many repair and calibration options for your analyzer. Contact the
nearest Hewlett-Packard sales or service office for information on options available for
your analyzer.
Summary of Contents for 8719ES
Page 6: ...vi ...
Page 10: ...Contents x Contents ...
Page 11: ...1 1 1 HP 8719 20 22ES Specifications and Characteristics ...
Page 60: ...1 50 HP 8719 20 22ES Specifications and Characteristics Instrument Specifications ...
Page 61: ...2 1 2 HP 8719 20 22ET Specifications and Characteristics ...
Page 98: ...2 38 HP 8719 20 22ET Specifications and Characteristics Instrument Specifications ...
Page 99: ...3 1 3 Front Rear Panel ...
Page 111: ...4 1 4 Menu Maps ...
Page 113: ...4 3 Menu Maps Menu Maps Figure 4 2 Menu Map for Copy ...
Page 114: ...4 4 Menu Maps Menu Maps Figure 4 3 Menu Map for Display ...
Page 115: ...4 5 Menu Maps Menu Maps Figure 4 4 Menu Map for Format Figure 4 5 Menu Map for Local ...
Page 116: ...4 6 Menu Maps Menu Maps Figure 4 6 Menu Map for Marker Marker Fctn and Marker Search ...
Page 118: ...4 8 Menu Maps Menu Maps Figure 4 9 Menu Map for Power and Sweep Setup ET only ...
Page 119: ...4 9 Menu Maps Menu Maps Figure 4 10 Menu Map for Power and Sweep Setup ES only ...
Page 120: ...4 10 Menu Maps Menu Maps Figure 4 11 Menu Map for Preset ...
Page 121: ...4 11 Menu Maps Menu Maps Figure 4 12 Menu Map for Save Recall ...
Page 122: ...4 12 Menu Maps Menu Maps Figure 4 13 Menu Map for Scale Ref ...
Page 128: ...5 1 5 Hardkey Softkey Reference ...
Page 194: ...6 1 6 Error Messages ...
Page 222: ...7 1 7 Options and Accessories ...
Page 234: ...8 1 8 Preset State and Memory Allocation ...
Page 253: ...8 20 Preset State and Memory Allocation Memory Allocation ...
Page 254: ...9 1 9 Understanding the CITIfile Data Format ...
Page 269: ...9 16 Understanding the CITIfile Data Format Useful Calculations ...
Page 270: ...10 1 10 Determining System Measurement Uncertainties ...
Page 281: ...10 12 Determining System Measurement Uncertainties Measurement Uncertainty Equations ...