![HP 8719ES Reference Manual Download Page 158](http://html.mh-extra.com/html/hp/8719es/8719es_reference-manual_161529158.webp)
5-31
Hardkey/Softkey Reference
Analyzer Functions
displays the current value of the loop counter and allows
you to change the value of the loop counter. Enter any
number from 0 to 32767 and terminate with the
key.
The default value of the counter is zero. This command
should be placed in a sequence that is separate from the
measurement sequence. For this reason: the measurement
sequence containing a loop decision command must call
itself in order to function. The
command must be in a separate sequence or the counter
value would always be reset to the initial value.
(Sequence Filenaming)
inserts the string "[LOOP]" into the file name.
accepts a power loss value for a segment in the power
meter cal power loss list. This value, for example, could be
the difference (in dB) between the coupled arm and
through arm of a directional coupler.
presents the power loss/sensor lists menu. This menu
performs two functions:
• Corrects coupled-arm power loss when a directional
coupler is used to sample the RF output.
• Allows calibration factor data to be entered for one or
two power sensors.
Each function provides up to 12 separate frequency points,
called segments, at which the user may enter a different
power loss or calibration factor. The instrument
interpolates between the selected points. Two power
sensor lists are provided because no single power sensor
can cover the frequency range of the network analyzer.
(Option 010 only) sets the transform to low pass impulse
mode, which simulates the time domain response to an
impulse input.
(Option 010 only) sets the transform to low pass step
mode, which simulates the time domain response to a step
input.
sets the lower limit value for the start of the segment in a
limit line list. If an upper limit is specified, a lower limit
must also be defined. If no lower limit is required for a
particular measurement, force the lower limit value out of
range (for example
−
500 dB).
mega/micro (10
6
/ 10
−
6
)
LOOP COUNTER
x1
LOOP COUNTER
LOOP COUNTER
LOSS
LOSS/SENSR LISTS
LOW PASS IMPULSE
LOW PASS STEP
LOWER LIMIT
M/
µ
Summary of Contents for 8719ES
Page 6: ...vi ...
Page 10: ...Contents x Contents ...
Page 11: ...1 1 1 HP 8719 20 22ES Specifications and Characteristics ...
Page 60: ...1 50 HP 8719 20 22ES Specifications and Characteristics Instrument Specifications ...
Page 61: ...2 1 2 HP 8719 20 22ET Specifications and Characteristics ...
Page 98: ...2 38 HP 8719 20 22ET Specifications and Characteristics Instrument Specifications ...
Page 99: ...3 1 3 Front Rear Panel ...
Page 111: ...4 1 4 Menu Maps ...
Page 113: ...4 3 Menu Maps Menu Maps Figure 4 2 Menu Map for Copy ...
Page 114: ...4 4 Menu Maps Menu Maps Figure 4 3 Menu Map for Display ...
Page 115: ...4 5 Menu Maps Menu Maps Figure 4 4 Menu Map for Format Figure 4 5 Menu Map for Local ...
Page 116: ...4 6 Menu Maps Menu Maps Figure 4 6 Menu Map for Marker Marker Fctn and Marker Search ...
Page 118: ...4 8 Menu Maps Menu Maps Figure 4 9 Menu Map for Power and Sweep Setup ET only ...
Page 119: ...4 9 Menu Maps Menu Maps Figure 4 10 Menu Map for Power and Sweep Setup ES only ...
Page 120: ...4 10 Menu Maps Menu Maps Figure 4 11 Menu Map for Preset ...
Page 121: ...4 11 Menu Maps Menu Maps Figure 4 12 Menu Map for Save Recall ...
Page 122: ...4 12 Menu Maps Menu Maps Figure 4 13 Menu Map for Scale Ref ...
Page 128: ...5 1 5 Hardkey Softkey Reference ...
Page 194: ...6 1 6 Error Messages ...
Page 222: ...7 1 7 Options and Accessories ...
Page 234: ...8 1 8 Preset State and Memory Allocation ...
Page 253: ...8 20 Preset State and Memory Allocation Memory Allocation ...
Page 254: ...9 1 9 Understanding the CITIfile Data Format ...
Page 269: ...9 16 Understanding the CITIfile Data Format Useful Calculations ...
Page 270: ...10 1 10 Determining System Measurement Uncertainties ...
Page 281: ...10 12 Determining System Measurement Uncertainties Measurement Uncertainty Equations ...