Front/Rear Panel
Analyzer Display
C2 (ES)
Full two-port error-correction is on and the reverse sweep
is not updated each sweep
Any one of the following causes the reverse sweep not to be
updated each sweep:
• the instrument uses a mechanical switch, for example
Options 85 and 007.
• different channel power ranges (PORT POWER
UNCOUPLED) which puts the test set switch in HOLD
mode except Option 400 (dual step attenuators).
• the user manually puts the test set switch in HOLD
mode (TESTSET SW 0 or >1).
Del
Electrical delay has been added or subtracted, or port
extensions are active. (See the “Operatiing Concepts”
chapter of the user’s guide and the
key in
Chapter 5 , “Hardkey/Softkey Reference.”
ext
Waiting for an external trigger.
Ofs (ES)
Frequency offset mode is on. (See the “Making Mixer
Measurements” chapter in the user’s guide.)
Of? (ES)
Frequency offset mode error, the IF frequency is not
within 10 MHz of expected frequency. LO inaccuracy is the
most likely cause. (See the “Making Mixer Measurements”
chapter in the user’s guide.)
Gat
Gating is on (time domain Option 010 only). (For time
domain measurement procedures and theory, refer to the
"Making Time Domain Measurements" chapter of the
user’s guide.)
Hld
Hold sweep. (See
in
man
Waiting for manual trigger.
PC
Power meter calibration is on. (For power meter
calibration procedures, refer to the "Calibrating for
Increased Measurement Accuracy" chapter of the user’s
guide.)
PC?
The analyzer's source could not be set to the desired level,
following a power meter calibration. (For power meter
calibration procedures, refer to the "Calibrating for
Increased Measurement Accuracy" chapter in the user’s
guide.)
Scale Ref
HOLD
Summary of Contents for 8719ES
Page 6: ...vi ...
Page 10: ...Contents x Contents ...
Page 11: ...1 1 1 HP 8719 20 22ES Specifications and Characteristics ...
Page 60: ...1 50 HP 8719 20 22ES Specifications and Characteristics Instrument Specifications ...
Page 61: ...2 1 2 HP 8719 20 22ET Specifications and Characteristics ...
Page 98: ...2 38 HP 8719 20 22ET Specifications and Characteristics Instrument Specifications ...
Page 99: ...3 1 3 Front Rear Panel ...
Page 111: ...4 1 4 Menu Maps ...
Page 113: ...4 3 Menu Maps Menu Maps Figure 4 2 Menu Map for Copy ...
Page 114: ...4 4 Menu Maps Menu Maps Figure 4 3 Menu Map for Display ...
Page 115: ...4 5 Menu Maps Menu Maps Figure 4 4 Menu Map for Format Figure 4 5 Menu Map for Local ...
Page 116: ...4 6 Menu Maps Menu Maps Figure 4 6 Menu Map for Marker Marker Fctn and Marker Search ...
Page 118: ...4 8 Menu Maps Menu Maps Figure 4 9 Menu Map for Power and Sweep Setup ET only ...
Page 119: ...4 9 Menu Maps Menu Maps Figure 4 10 Menu Map for Power and Sweep Setup ES only ...
Page 120: ...4 10 Menu Maps Menu Maps Figure 4 11 Menu Map for Preset ...
Page 121: ...4 11 Menu Maps Menu Maps Figure 4 12 Menu Map for Save Recall ...
Page 122: ...4 12 Menu Maps Menu Maps Figure 4 13 Menu Map for Scale Ref ...
Page 128: ...5 1 5 Hardkey Softkey Reference ...
Page 194: ...6 1 6 Error Messages ...
Page 222: ...7 1 7 Options and Accessories ...
Page 234: ...8 1 8 Preset State and Memory Allocation ...
Page 253: ...8 20 Preset State and Memory Allocation Memory Allocation ...
Page 254: ...9 1 9 Understanding the CITIfile Data Format ...
Page 269: ...9 16 Understanding the CITIfile Data Format Useful Calculations ...
Page 270: ...10 1 10 Determining System Measurement Uncertainties ...
Page 281: ...10 12 Determining System Measurement Uncertainties Measurement Uncertainty Equations ...